Specifications |
PART NUMBER: 47-1560
APPLICATION: beta,
low energy gamma radiation detection
RADIATION DETECTOR: scintillator, 1 in. (2.54 cm) diameter x 1 mm thick
Nal adhered to a 0.001 in. (0.03 cm) thick plastic scintillator
WINDOW: 3.4 mg/cm²
WINDOW AREA: 5.1 cm² active and open
EFFICIENCY(4π):
38% - 125I;
19% - 129I;
5% - 14C; 28% - 32P
NON-UNIFORMITY: < 10%
ENERGY RESPONSE: energy dependent (see graph below)
BACKGROUND (10 µR/hr):
450 cpm
COMPATIBLE INSTRUMENTS: general purpose radiation survey meters,
ratemeters and scalers
PHOTOMULTIPLIER TUBE: 1.5 in. (3.8cm) diameter
OPERATING VOLTAGE: typically 500 to 1200 volts
CONNECTOR: series "C"
(others
available)
SIZE: 2 x 7.3 in. (5.1 x 18.5 cm) (Dia x L), with connector cap
WEIGHT: 1 lb (0.5 kg)
CE CERTIFIED: yes

|
Options & Accessories |
|
|
|
|
![]() |
|
|
Sample Holder |
Sample Holder |
Sample Holder |
Sample Planchets |
|
Technical Information |
|
CE
Certificate |
Energy Response |
|
Ludlum Resources |